Design and calibration of a 60-GHz personal exposimeter for exposure assessment in specular and diffuse environments - Université de Rennes Accéder directement au contenu
Communication Dans Un Congrès Année : 2017

Design and calibration of a 60-GHz personal exposimeter for exposure assessment in specular and diffuse environments

R. Aminzadeh
  • Fonction : Auteur
A. Thielens
  • Fonction : Auteur
H. Li
  • Fonction : Auteur
G. Torfs
  • Fonction : Auteur
J. Bauwelinck
  • Fonction : Auteur
L. Martens
  • Fonction : Auteur
W. Joseph
  • Fonction : Auteur

Résumé

For the first time, a personal exposimeter (PE) for 60 GHz radiation measurements is presented. The PE is designed based on numerical simulations, on-phantom and on-body calibration measurements under specular and diffuse exposure to determine the antenna aperture and measurement uncertainty of the PE. A 95% prediction interval of 6.6 dB is obtained for the PE's response using a combination of three nodes with different polarizations. The response of the PE is in the range of 0.7 to 1.2 for specular and diffuse fields. © 2017 EAI.

Dates et versions

hal-01637409 , version 1 (17-11-2017)

Identifiants

Citer

R. Aminzadeh, A. Thielens, A.K. Fall, H. Li, C. Leduc, et al.. Design and calibration of a 60-GHz personal exposimeter for exposure assessment in specular and diffuse environments. 11th International Conference on Body Area Networks, BODYNETS 2016, Dec 2016, Turin, Italy. ⟨10.4108/eai.15-12-2016.2267789⟩. ⟨hal-01637409⟩