Impact of early complications on outcomes in patients with implantable cardioverter-defibrillator for primary prevention - Université de Rennes Accéder directement au contenu
Article Dans Une Revue Heart Rhythm Année : 2016

Impact of early complications on outcomes in patients with implantable cardioverter-defibrillator for primary prevention

Eloi Marijon
Jean-Claude Deharo
Olivier Piot
  • Fonction : Auteur correspondant
  • PersonId : 948860

Connectez-vous pour contacter l'auteur

Résumé

Background - The lifesaving benefit of implantable cardioverter-defibrillators (ICDs) has been demonstrated. Their use has increased considerably in the past decade, but related complications have become a major concern. Objective - The purpose of this study was to assess the incidence and effect on outcomes of early (≤30 days) complications after ICD implantation for primary prevention in a large French population. Methods - We analyzed data from 5539 patients from the multicenter French DAI-PP (Défibrillateur Automatique Implantable-Prévention Primaire) registry (2002-2012) who had coronary artery disease or dilated cardiomyopathy and were implanted with an ICD for primary prevention. Results - Overall, early complications occurred in 707 patients (13.5%), mainly related to lead dislodgment or hematoma (57%). Independent factors associated with occurrence of early complications were severe renal impairment (odds ratio [OR] 1.66, 95% confidence interval [CI] 1.17-2.37, P = .02), age ≥75 years (OR 1.01, 95% CI 1.00-1.02, P = .03), cardiac resynchronization therapy (OR 1.58, 95% CI 1.16-2.17, P = .01), and anticoagulant therapy (OR 1.28, 95% CI 1.02-1.61, P = .03). During a mean ± SD follow-up of 3.1 ± 2.3 years, 824 (15.8%) patients experienced ≥1 late complication (>30 days), and 782 (14.9%) patients died. After adjustment, early complications remained associated with occurrence of late complications (OR 2.15, 95% CI 1.73-2.66, P < .0001) and mortality (OR 1.70, 95% CI 1.34-2.17, P = .003). Conclusion - Early complications are common after ICD implantation for primary prevention, occurring in 1 in 7 patients, and are associated with an increased risk of late complications and overall mortality. Further studies are needed to investigate the underlying mechanisms of such associations.
Fichier principal
Vignette du fichier
Impact of Early Complications on Outcomes postprint.pdf (1.51 Mo) Télécharger le fichier
Origine : Fichiers produits par l'(les) auteur(s)

Dates et versions

hal-01272165 , version 1 (10-02-2016)

Identifiants

Citer

Maria Soledad Ascoeta, Eloi Marijon, Pascal Defaye, Didier Klug, Frankie Beganton, et al.. Impact of early complications on outcomes in patients with implantable cardioverter-defibrillator for primary prevention. Heart Rhythm, 2016, 13 (5), pp.1045-1051. ⟨10.1016/j.hrthm.2015.12.046⟩. ⟨hal-01272165⟩
394 Consultations
269 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More