Theoretical analysis of electromagnetic exposure to wireless charging systems for deep-body implantable devices - Université de Rennes Accéder directement au contenu
Communication Dans Un Congrès Année : 2021

Theoretical analysis of electromagnetic exposure to wireless charging systems for deep-body implantable devices

Résumé

Wireless Power Transfer (WPT) techniques allow significant miniaturisation of implanted medical devices. However, the electromagnetic exposure of the body tissues must be accurately accessed to ensure safe operation. This study, therefore, analyses the exposure induced by the magnetic-and electric-type antenna transmitters over the frequency range of 100-MHz-5-GHz. The normalised power absorption, specific absorption rate (SAR) and WPT efficiency are analysed and presented. The results demonstrate the existence of an optimal operating frequency at which the maximum efficiency, minimum power absorption, and low SAR are simultaneously observed.
Fichier principal
Vignette du fichier
Paper_BioEM_2021.pdf (2.09 Mo) Télécharger le fichier
Origine : Fichiers produits par l'(les) auteur(s)

Dates et versions

hal-03384083 , version 2 (18-10-2021)
hal-03384083 , version 1 (11-11-2023)

Identifiants

  • HAL Id : hal-03384083 , version 1

Citer

Icaro V. Soares, Ronan Sauleau, Denys Nikolayev. Theoretical analysis of electromagnetic exposure to wireless charging systems for deep-body implantable devices. BioEM 2021 (The Joint Annual Meeting of The Bioelectromagnetics Society and the European BioElectromagnetics Association), Sep 2021, Gand, Belgium. ⟨hal-03384083v1⟩
91 Consultations
34 Téléchargements

Partager

Gmail Facebook X LinkedIn More