Study of thrombin generation with St Genesia to evaluate xaban pharmacodynamics: Analytical performances over 18 months

Résumé

INTRODUCTION: ST Genesia is a new automated system enabling quantitative standardized evaluation of thrombin generation (TG), for example, in patients receiving anti-Xa direct inhibitors (xabans). Data on its analytical performances are scarce. METHODS: Over an 18-month period, repeatability, reproducibility, and accuracy were assessed using STG-ThromboScreen (without or with thrombomodulin) or STG-DrugScreen reagents (corresponding to intermediate/high tissue-factor concentration, respectively), and controls. Furthermore, reproducibility was assessed using commercialized lyophilized and frozen normal pooled plasmas. Rivaroxaban and apixaban impacts on TG parameters were assessed using spiking experiments. Finally, a comparison with the Calibrated Automated Thrombogram method (CAT) (PPP reagent) was performed using plasma from healthy volunteers enrolled in the DRIVING-studyNCT 01627665) before and after rivaroxaban intake. RESULTS: For all dedicated quality control (QC) levels, inter-series coefficients of variations (CV) were <7% for temporal TG parameters, peak height (PH), and endogenous thrombin potential (ETP), whether results were normalized with a dedicated reference plasma STG-RefPlasma or not. Noteworthy, STG-RefPlasma used for normalization displayed substantially high PH and ETP. Mean biases between the observed and manufacturer’s assigned QC values were mostly <7%. Both rivaroxaban/apixaban plasma concentrations were significantly associated with TG parameters. Finally, Bland-Altman plots showed a good agreement between ST Genesia-STG-ThromboScreen and CAT method within the explored range of values, although biases could be observed (PH: 16.4 ± 13.2%, ETP: 17.8 ± 11.9%). CONCLUSION: ST Genesia(®) enables the reliable measurement of TG parameters in both in vitro and ex vivo xaban plasma samples using either STG-ThromboScreen or STG-DrugScreen according to xaban concentrations. The use of reference plasma, despite not completely reflecting a normal pooled plasma behavior, likely improves standardization and inter-laboratory comparisons.
Fichier non déposé

Dates et versions

hal-03102064 , version 1 (07-01-2021)

Identifiants

Citer

Geoffrey Foulon-Pinto, Georges Jourdi, Julien Perrin, Johan Abdoul, Guillaume Paris, et al.. Study of thrombin generation with St Genesia to evaluate xaban pharmacodynamics: Analytical performances over 18 months. International Journal of Laboratory Hematology, 2021, 43 (4), pp.821-830. ⟨10.1111/ijlh.13443⟩. ⟨hal-03102064⟩
88 Consultations
0 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More