ViP-CSI Virtual Phantom Chemical Shift Imaging
Résumé
To apply the Virtual Phantom (ViP) method for generating reference signals, to the chemical shift imaging (CSI) technique. The ViP method, previously used for generating reference images in magnetic resonance imaging (MRI), was here extended to generate reference peaks in the MR spectra acquired with a 2D-CSI sequence. Theoretical analysis was carried out to design the ViP signal. ViP-2D-CSI experiments were performed on a 4.7 Tesla MR scanner. Data analysis was carried out with the jMRUI software to fit the ViP peaks to determine their amplitude and frequency. Based on the theoretical analysis, it was found that the ViP peak could be generated by transmitting the ViP signal only in one repetition time of the 2D-CSI sequence. The amplitude and frequency of the ViP peak could be precisely adjusted and fine-tuned. The ViP peak was uniform over all CSI voxels, both in amplitude and in frequency. Furthermore, a good stability of the ViP reference amplitude and frequency was observed. The ViP method provides a means to generate a reference MR peak in 2D-CSI experiments. This could be of interest for signal quantification in CSI experiments.
Domaines
Imagerie médicale
Fichier principal
Virtual Phantom Chemical Shift Imaging_accepted.pdf (13.77 Mo)
Télécharger le fichier
Origine | Fichiers produits par l'(les) auteur(s) |
---|